6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1435
Warning limits for E(0) (for ISE electrodes only)
on | off (Default value: off)
If this option is activated, then the values for E(0) will be shown in
orange if the limits are exceeded.
Lower limit
Lower warning limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-15 mV
Upper limit
Upper warning limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
15 mV
Intervention limits for E(0) (for ISE electrodes only)
on | off (Default value: off)
If this option is activated, then the values for E(0) will be shown in red if
the limits are exceeded.
Lower limit
Lower intervention limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-30 mV
Upper limit
Upper intervention limit for E(0).
Range
-2.147E+9 ... 2.147E+9 mV
Default value
30 mV
Warning limits for cell constant (for conductivity sensors
only)
on | off (Default value: off)
If this option is activated, then the values for the cell constant will be
shown in orange if the limits are exceeded.
Lower limit
Lower warning limit for the cell constant.
Range
0.001 ... 500 /cm
Default value
2.0 /cm