4 dos u - stop conditions, Stop conditions – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1203
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1187
5.6.6.3.5.4
DOS U - Stop conditions
Tab: Method
▶ DOS U ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
The conditions for stopping the dosing are defined on this tab. If more
than one stop condition is in effect at the same time, then the criterion
which is fulfilled first will stop the dosing.
Stop volume
Stops when the given volume has been added after the start of the dos-
ing. The stop volume should be adapted to suit the measuring vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100.000 mL
Selection
off
Titrino
Range
0 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Range
0 ... 999999 s
Selection
off
Default value
off
off
Stops when the preset time has elapsed since the start of the dosing.
Filling rate
Rate at which the dosing cylinder is to be filled after dosing. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used. If the entered filling rate is too high for the selected
dosing device, it will automatically be reduced during dosing to the largest
possible value.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min