5 special beaker – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1495
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1479
Special beaker
Number of the special beaker for the selected rack.
Rack position
Number of the rack position for the special beaker.
Work position Tower 1
Work position for the special beaker at Tower 1.
Work position Tower 2
Work position for the special beaker at Tower 2.
Beaker radius
Radius of the special beaker.
Beaker sensor
Shows whether and which beaker sensor is to be used for the special
beaker.
[Edit]
Opens the dialog window Special beaker # (see Chapter 6.12.3.5, page
1479) for editing the data of the selected special beaker.
6.12.3.5
Special beaker
Dialog window: Configuration
▶ Rack data ▶ [Edit] ▶ Properties... ▶ Rack
data
▶ Special beaker ▶ [Edit] ▶ Special beaker #
The parameters for the special beaker selected in the table of special beak-
ers can be defined in the dialog window Special beaker #.
Rack position
Number of rack position for selected special beaker.
Range
0 ... n (n is rack-dependent)
Note
Special beakers should preferably be set at high rack positions so that
sample series can start from rack position 1. Rack positions which are
defined as special beakers are no longer available as sample positions.
They will be skipped during an automatic movement to a sample posi-
tion.
Work position Tower 1
Work position of the selected special beaker at Tower 1.