5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1150
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Temperature
Temperature
The temperature at which calibration is carried out can be entered man-
ually. If a temperature sensor is connected and the Temperature meas-
urement is set on the General/Hardware tab under Sensor to auto-
matic or continuous, then the temperature will be measured continu-
ously. The value is used for temperature correction with the pH measure-
ment.
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
5.6.6
Dosing commands
5.6.6.1
Dosing commands - Overview
Menu item: Method
▶ Insert ▶ New command... ▶ Dosing
Commands for Dosing with dosing units and exchange units.
The following dosing commands can be selected:
■
ADD
Adding a predefined volume.
■
DOS pH
Controlled dosing with monitoring of measured value pH and tempera-
ture
■
DOS U
Controlled dosing with monitoring of measured value U and tempera-
ture
■
LQH
Extensive dosing possibilities with a Dosino.
■
PREP
Preparing a exchange or dosing unit.
■
EMPTY
Emptying a dosing unit.
■
RLS DOS
Releasing a dosing unit.
5.6.6.2
ADD
5.6.6.2.1
ADD - Overview
Dialog window: Method
▶ ADD ▶ Properties... ▶ ADD - 'Command name'
Command for dosing a defined volume of a solution with an exchange
or dosing unit.
Devices
This command can be executed with the following devices: