4 kfc - control parameters – Metrohm tiamo 2.3 (process analysis) User Manual
Page 847
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
831
Range
0 ... 999999 s
Default value
0 s
5.6.3.6.4
KFC - Control parameters
Tab: Method
▶ KFC ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Input
25 characters
Endpoint
EP with
Measured value for the endpoint.
756, 831
Range
-2000.0 ... 2000.0 mV
Default value
50.0 mV
851, 852
Range
-1250.0 ... 1250.0 mV
Default value
50.0 mV
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.
Selection
slow | optimal | fast | User
Default value
optimal
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical, fast titrations.
User
User-defined settings of the individual titration parameters which affect
the titration rate.