Metrohm tiamo 2.3 (process analysis) User Manual
Page 831
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
815
Stop drift
Titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.
Range
1 ... 999 µL/min
Default value
20 µL/min
Relative stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion = rel. drift.
Titrando
Range
1 ... 999 µL/min
Default value
10 µL/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time. The selection inf. stands for infin-
ite.
Titrando, 855
Range
0 ... 999 s
Default value
10 s
Titrino
Range
0 ... 999 s
Default value
10 s
Selection
inf.
Stop time
The titration is stopped if the absolute stop time defined here has elapsed
since the start of the titration. This parameter is only shown for Stop cri-
terion = time and Delay time = inf.
Titrino
Range
1 ... 999999 s