5 signatures, Electronic signature, Signatures – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1350
6.2 Administration
■■■■■■■■■■■■■■■■■■■■■■
1334
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
Modifications
Comment on modification of methods
If this check box is activated then each time that a method is modified a
modification reason and a modification comment must be entered; these
are saved in the method and shown in the method history. The reason
and comments are also recorded in the Audit Trail.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
Comment on modification of determinations
If this check box is activated, then a reason for a change and a change
comment must be entered with each change to a determination which
will then be saved in the determination and displayed in the database in
the subwindow Information on the tab Determination. Reason and
comment will also be recorded in the Audit Trail.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
Comment on modification of sample data (live)
If this check box is activated then each time that a determination is
modified a modification reason and a modification comment must be
entered; these are saved in the determination and shown in the database
in the subwindow Information on the tab Sample. The reason and
comments are also recorded in the Audit Trail.
Selection
on | off
Default value
off
FDA default: on (can no longer be edited)
6.2.2.5
Signatures
Tab: Configuration
▶ Tools ▶ Security settings ▶ Signatures
On the tab Signatures, the parameters for the electronic signature can
be configured.
Inactivity delay
If this check box is activated then the dialog window for signing will be
automatically closed when the entered time limit has expired.