5 usable electrode – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1377
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1361
Max. slope
Maximum permitted slope of the pH electrode.
Range
-999.9 ... 999.9 %
Default value
102.0 %
Max. response time
Maximum permitted response time of the pH electrode. The voltage
measured after three minutes in stirred solution is used as a comparison
value for determining the response time. The response time is the time
after which the measured voltage has reached this comparison value
except for ±1 mV.
Range
0 ... 999 s
Default value
50 s
Min. offset potential
Minimum permitted offset voltage (voltage at pH = 7.0). This value applies
for all evaluations.
Range
-999 ... 999 mV
Default value
-15 mV
Max. offset potential
Maximum permitted offset voltage (voltage at pH = 7.0). This value
applies for all evaluations.
Range
-999 ... 999 mV
Default value
15 mV
6.3.3.6.5
Usable electrode
Tab: Configuration
▶ Tools ▶ Templates ▶ Electrode types... ▶ [Properties] ▶ Usa-
ble electrode
Max. streaming potential
Maximum permitted flow potential (difference of the measured voltages
in stirred and non-stirred solution).
Range
0.0 ... 999.9 mV
Default value
4.0 mV
Max. drift
Maximum permitted drift in stirred solution.
Range
0.1 ... 9.9 mV/min
Default value
3.0 mV/min