Metrohm tiamo 2.3 (process analysis) User Manual
Page 743
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
727
off
The titration will not be stopped. In older instructions the Delay time
was usually defined as the Stop criterion.
Stop drift
Titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.
Range
1 ... 999 µL/min
Default value
20 µL/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time. The selection inf. stands for infin-
ite.
Titrando, 855
Range
0 ... 999 s
Default value
10 s
Titrino
Range
0 ... 999 s
Default value
10 s
Selection
inf.
Stop time
The titration is stopped if the stop time defined here has elapsed since the
start of the titration. This parameter is only shown for Stop criterion =
time and Delay time = inf.
Titrino
Range
0 ... 999999 s
Selection
off
Default value
off
Endpoint 2
Note
Is activated only for EP1 with ≠ off.
on | off (Default value: off)
Switching endpoint 2 on/off. If this is deactivated all following parameters
are not visible.