Metrohm tiamo 2.3 (process analysis) User Manual
Page 882
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
866
■■■■■■■■
tiamo 2.3 (for Process Analysis)
855
Selection
1 | 2 | 3
Default value
2
Solution
Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method run to see whether the
correct solution has been set on the selected dosing device and whether
the dosing device type is correct. With non-intelligent exchange or dosing
units, only the cylinder volume is checked.
Titrando, 855
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
Filling rate
Rate at which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during dos-
ing to the largest possible value (see Chapter 5.6.3.8.5, page 859).
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Sensor
Measuring input
Selection of the measuring input to which the sensor is connected.
Titrando
Selection
1 | 2
Default value
1
855
Selection
1
Default value
1