6 met ipol - potentiometric evaluation, Potentiometric evaluation – Metrohm tiamo 2.3 (process analysis) User Manual
Page 650
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
634
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.3.4.6
MET Ipol - Potentiometric evaluation
Tab: Method
▶ MET Ipol ▶ Properties... ▶ Potentiometric evaluation
Command name
Name of the command.
Input
25 characters
Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition are defined on this tab. One of
the following three evaluation methods can be chosen here:
Potentiometric evaluation
Selection
Evaluation without window | Evaluation with
measured value window (Upol) | Evaluation
with volume window (mL)
Default value
Evaluation without window
Evaluation without window
With this option set, the two parameters EP criterion and EP recog-
nition will be applied across the entire range of the titration curve.
Evaluation with measured value window (Upol)
With this option set, up to 9 regions (windows) can be defined on the
measured value axis. Only those equivalence points which lie within
these windows and additionally meet the parameters defined for each
window will be recognized. Per window only one EP will be recog-
nized. The defined windows with their parameters are shown in the
window table and can be edited with the following buttons: