3 analog in - stop criteria – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1315
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1299
5.6.11.5.3
ANALOG IN - Stop criteria
Tab: Method
▶ ANALOG IN ▶ Stop criteria
Command name
Name of the command.
Input
25 characters
Stop criteria
Definition of possible stop criteria. The first stop criterion to be fulfilled will
stop the command.
Maximum run time
on | off (Default value: off)
Check box for switching on/off the stop criterion Maximum run time. If
the maximum run time is reached, then the command will be stopped.
Range
0 ... 999.9
Default value
0
Selection
min | s
Default value
min
Note
The run time is interrupted if the method is stopped with [HOLD].
Condition
on | off (Default value: off)
Check box for switching on/off the stop criterion Condition.
Expression for the condition(s) which can be created or edited with the
formula editor after pressing on
or double-clicking in the text field
The condition is monitored on a continuous basis; if it is fulfilled (evalua-
tion of the formula gives the result 1 = true), then the command will be
interrupted.
Input
1000 characters
Default value
empty