6 met upol - potentiometric evaluation, Potentiometric evaluation – Metrohm tiamo 2.3 (process analysis) User Manual
Page 676
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
660
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Volume after EP
When the number of equivalence points defined under Stop EP has been
found, this volume will be added. In this way you can see the curve shape
after the equivalence point is reached.
Titrando, 855
Range
0.01000 ... 9999.99 mL
Selection
off
Default value
off
Stop time
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.
Titrando, 855
Range
0 ... 999999 s
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.3.5.6
MET Upol - Potentiometric evaluation
Tab: Method
▶ MET Upol ▶ Properties... ▶ Potentiometric evaluation
Command name
Name of the command.
Input
25 characters
Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition are defined on this tab. One of
the following three evaluation methods can be chosen here: