2 curves - measuring point list, 3 curves - monitoring report, Curves - measuring point list – Metrohm tiamo 2.3 (process analysis) User Manual
Page 396: Curves - monitoring report
4.8 Subwindow Curves
■■■■■■■■■■■■■■■■■■■■■■
380
■■■■■■■■
tiamo 2.3 (for Process Analysis)
¯Dos.r.
Upper limit value for dosing rate monitoring infringed.
_Temp.
Lower limit value for temperature monitoring infringed.
¯Temp.
Upper limit value for temperature monitoring infringed.
*
The dosing has been paused.
4.8.2
Curves - Measuring point list
Subwindow: Database
▶ Curves
With the context-sensitive menu item Measuring point list… in the
curve window the dialog window Measuring point list - 'Command
name.#' opens for showing the measuring point list for the selected
curve.
The display of the measuring points in the table can be defined per win-
dow and per command type in the Properties - Curve #.
4.8.3
Curves - Monitoring report
Subwindow: Database
▶ Curves
If the method contains commands of the type STAT or DOS, then with
the context-sensitive menu item Monitoring report… in the curve win-
dow the dialog window Monitoring report - 'Command name' can
be opened. In this window information on limit violations occurred are
displayed for each monitoring switched on.
The following columns are displayed:
Time [s]
Time in s until limit violation.
Limit
Selection
lower | upper | ok
lower
Lower limit value exceeded.
upper
Upper limit value exceeded.
ok
Measured value within the limits again.
[pH]/[U]
Limit for the measured value.