Metrohm tiamo 2.3 (process analysis) User Manual
Page 835
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
819
Drift correction
If this parameter is set to automatic or manual the drift correction ( =
Drift value × Time for drift correction) caused by empty consumption
during titration is deducted from the total consumption. The Time for
drift correction is the period between end of conditioning and end of
titration.
Selection
automatic | manual | off
Default value
off
automatic
The drift lastly determined during conditioning will be used as drift
value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 99.9 µL/min
Default value
0.0 µL/min
Stop volume
Determination stops when the specified volume has been dosed since the
start of the conditioning. The stop volume should be adjusted to the size
of the titration cell in order to prevent any overflow.
Titrando
Range
0.000000 ... 9999.99 mL
Default value
20.00 mL
Selection
off
Default value
off
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Titrando
Range
1 ... 999999 s
Selection
off
Default value
off