5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
743
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos and 855.
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the