4 properties - limit value, Limit values – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1461
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1445
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the validity of the calibration has expired will be
automatically saved with the determination.
Display message
A message is displayed and you can select whether the run is to be
continued or canceled. If the run is continued then the message that
the validity of the calibration has expired will be saved automatically
with the determination.
Cancel determination
The running determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].
6.8.3.4
Properties - Limit value
Tab: Configuration
▶ Colorimetric sensors ▶ Edit ▶ Properties... ▶ Limits
Monitoring number of outliers
on | off (Default value: off)
If this check box is activated, then the number of outliers during a new
calibration is monitored. If this number exceeds the defined limit value,
then a corresponding message will appear and one can decide whether
the calibration should nevertheless be applied or if the previous calibration
data should be retained.
Upper limit
Maximum number of permitted outliers.
Only editable for Monitoring number of outliers = on, otherwise inac-
tive.
Range
0 ... 9
Default value
1
Monitoring coefficient of determination
on | off (Default value: off)
If this check box is activated, then the coefficient of determination (r
2
) cal-
culated from the calibration data will be monitored after a new calibra-
tion. If this number lies outside the defined range, then a corresponding
message will appear and one can decide whether the calibration should
nevertheless be applied or if the previous calibration data should be
retained.