Metrohm tiamo 2.3 (process analysis) User Manual
Page 1488
6.11 Subwindow Sample solutions (TC conductivity)
■■■■■■■■■■■■■■■■■■■■■■
1472
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Message
The message options can only be edited if monitoring is switched on.
Message by E-mail
on | off (Default value: off)
The message is additionally sent to the e-mail address defined under [E-
mail] if this option is activated. The message is sent in text format.
[E-mail]
[E-Mail] opens the Send E-mail window (see Chapter 2.6.1, page 92).
Acoustic signal
on | off (Default value: off)
If this option is activated, an acoustic signal will be emitted additionally to
the message defined above.
Action
The actions can only be edited if monitoring is switched on.
If during monitoring it is found that the validity period has expired then
one of the following actions will be triggered automatically at start test.
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the validity period has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period has expired will be automatically saved with the
determination.
Cancel determination
The running determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].