Metrohm tiamo 2.3 (process analysis) User Manual
Page 1487
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1471
Assignment date
Display of the moment at which the data for the calculation of the tem-
perature coefficient was entered or determined. This value is entered
automatically by tiamo.
Assignment method
Name of the method with which the automatic value assignment took
place. If the value has been entered manually then manual will be
shown.
User
Short name of the user who was logged in during value assignment or
who entered the value manually. If work is not carried out using login,
then the user logged in under Windows will be entered automatically.
Measuring input
Type and serial number of the measuring input the conductivity sensor
was connected to.
Determination ID
Unambiguous identification for the determination with which the determi-
nation of the TC data was carried out.
The determination ID can be copied and pasted into another field.
Monitoring TC data
on | off (Default value: off)
If this option is activated, then the validity period of the data will be moni-
tored.
Validity
Validity period of the data in days. If a value is entered here, then the date
in the field Next assignment will be adapted automatically.
Range
0 ... 999 Days
Default value
999 Days
Next assignment
Date on which the next assignment is to be carried out. The date can be
selected by pressing
in the dialog window Select date (see Chapter
2.5.1, page 89). After a date has been entered, the value for the Validity
will be automatically adapted.
Selection
Date selection