5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
906
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Monitoring Temperature
Monitoring Temperature
on | off (Default value: off)
If this check box is activated, then the temperature will be monitored and
any exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the temperature. If the temperature falls below this limit,
the event Lower temperature limit exceeded is triggered.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
-20.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
-170.0 °C
Lower hysteresis
Lower hysteresis of the temperature. If the temperature once again
exceeds the lower limit by this hysteresis value, then the event Tempera-
ture lower limit ok is triggered.
Titrando, 855
Range
0.0 ... 150.0 °C
Default value
0.2 °C
Upper limit
Upper limit of the temperature. If the temperature exceeds this limit, the
event Temperature upper limit exceeded is triggered.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
150.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
500.0 °C
Upper hysteresis
Upper hysteresis of the temperature. If the temperature once again falls
below the upper limit by this hysteresis value, then the event Tempera-
ture upper limit ok is triggered.