Metrohm tiamo 2.3 (process analysis) User Manual
Page 853
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
837
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 99.9 µg/min
Default value
0.0 µg/min
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Range
1 ... 999999 s
Selection
off
Default value
off
Stabilizing time
Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.
Range
0 ... 999 s
Default value
0 s
Automatic start after sample addition time
on | off (Default value: on)
If this option is activated, then the titration will be started automatically
after the sample addition time.
Sample addition time
Waiting time between stop of conditioning and start of titration during
which the sample can be added.
Range
0 ... 99 s
Default value
0 s
Manual start after sample addition with [Continue]
on | off (Default value: off)
If this option is activated, then the titration will be started manually after
the sample addition by pressing the [Continue] button in the message
window.