6 kft ipol - stop conditions, Stop conditions – Metrohm tiamo 2.3 (process analysis) User Manual
Page 812
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
796
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Time interval measuring point
Time interval for entering a measuring point in the measuring point list.
Titrando
Range
0.1 ... 999999 s
Default value
2.0 s
Titrino
Range
0.08 ... 16200 s
Default value
2.00 s
5.6.3.5.2.6
KFT Ipol - Stop conditions
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.
Stop volume
Stop when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Titrando
Range
0.00000 ... 9999.99 mL
Default value
100.000 mL
Selection
off
Titrino
Range
0.00 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stop when the entered time has elapsed after the start of the titration
(including start conditions).
Titrando
Range
0 ... 999999 s
Selection
off
Default value
off