5 tet - stop conditions – Metrohm tiamo 2.3 (process analysis) User Manual
Page 934
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
918
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Damping until
Range from the start of the measurement in which the curve is addition-
ally damped. This parameter allows to minimize the influence of titration
artifacts (e.g. spikes), which occur at the beginning of the measurement
and can mistakenly be interpreted as endpoints. The damping functions as
a linear ramp, i.e. its influence is the highest at the beginning and then
decreases linearly to the defined volume.
Range
0.0 ... 50.0 mL
Default value
0.2 mL
5.6.3.9.5
TET - Stop conditions
Tab: Method
▶ TET ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.
Stop volume
Stop when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Range
0.01 ... 50.00 mL
Selection
off
Default value
off
Stop measured value
Stop when the measured value entered for a measuring point has been
reached since the start of the titration. Thereby, the direction of the meas-
urement curve (ascending or descending), which is determined after
recording the first 50 measuring points, is taken into account.
Range
0.0 ... 60.0 °C
Selection
off
Default value
off
Stop time
Stop when the entered time has elapsed after the start of the titration
(including start conditions).
Range
0 ... 999999 s