6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1443
Calibration temperature
Temperature at which the calibration is carried out. In addition, the display
shows whether the temperature was measured with a Pt1000 or an NTC
temperature sensor or entered manually.
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
Note
If the Calibration temperature has been entered manually, then the
following parameters will be reset to the default value or updated,
respectively:
■
Calibration temperature = manual
■
Calibration date is updated.
■
Calibration method = manual
■
Determination ID is deleted
■
User is updated
■
Audit Trail entry ensues
Calibration date
Display of date and time of the last calibration entered after each auto-
matic calibration or manual entry.
Calibration method
Display of the method name with which the last calibration has been car-
ried out. If the calibration data has been entered manually, then manual
will appear.
Determination ID
Identification of the determination.
The determination ID can be copied and pasted into another field.
User
Display of the short name of the user logged in during the calibration or
who entered the calibration data manually. If work is not carried out using
login, then the user logged in under Windows will be entered automati-
cally.
Calibration data monitoring
on | off (Default value: off)
If this check box is activated, then the validity of the calibration is moni-
tored.