5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1164
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Identifica-
tion
Description
.FIN
Command status; 1 = Command has been ended
at least once, 0, invalid (variable not available) =
Command has never been ended
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value × (1...3) for the last measuring point
of the measuring point list
.LP.DVT
dV/dt for the last measuring point in the measuring
point list (SET, KFT, STAT, DOS) or drift for the last
measuring point in in the measuring point list in
µg/min (KFC)
.LP.EXx
External value × (1...3) for the last measuring point of
the measuring point list
.LP.MEA
Measured value for the last measuring point in the unit
of the measured value of the measuring point list
.LP.TEM
Temperature for the last measuring point of the meas-
uring point list in °C
.LP.TIM
Time in s until the last measuring point in the measur-
ing point list is reached
.LP.VOL
Volume for the last measuring point in the measuring
point list in mL
.MR.MRC
Correlation coefficient for mean dosing rate for the
whole range
.MR.MRS
Standard deviation for mean dosing rate for the whole
range in mL/min
.MR.MRT
Mean dosing rate for the whole range in mL/min.
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in the measuring point list
.SLO
Electrode slope of the sensor used for the command (in
%)