2 analog in - general – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1314
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1298
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command has
never ended
.UNI
Mapping unit which is defined in the configuration for an
IO port.
.[VAL]
Value of the command variables (optional, i.e. Command
name.Port name' = Command name.Port
name.VAL'). The value returned corresponds to the status
of the IO port after completion of the command; 1 = acti-
vated, 0 = deactivated
5.6.11.5.2
ANALOG IN - General
Tab: Method
▶ ANALOG IN ▶ General
Command name
Name of the command.
Input
25 characters
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display of the device type.
Selection
IO Controller
Default value
IO Controller