5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
875
Stop volume
Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stop when the entered time has elapsed after the start conditions have
been fulfilled.
Range
0 ... 999999 s
Selection
off
Default value
off
since
Specifies when the stop time begins to be counted down (will not be dis-
played if the Stop time is disabled):
Selection
Start | Control point first reached | Last dosing
Default value
Start
Start
The stop time starts after the start volume has been dosed and after
the following pause time.
Control point first reached
The stop time starts as soon as the control point has been reached for
the first time.
Last dosing
The stop time starts after the last increment has been dosed, i. e. the
stop time is reset to zero again after each dosing step.
Stop rate
Stop when the current dosing rate is lower than the defined stop rate.
This parameter is not checked until 10 s after the start. The maximum stop
rate depends on the cylinder volume of the exchange unit or dosing unit
used.