5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
902
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.8.7.7
STAT U - Monitoring
Tab: Method
▶ STAT U ▶ Properties... ▶ Monitoring
Command name
Name of the command.
Input
25 characters
The following monitoring processes can be switched on and defined on
this tab:
Monitoring Measured value
Monitoring Measured value
on | off (Default value: off)
If this check box is activated, then the measured value will be monitored
and any exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the measured value. If the measured value falls below this
limit, the event Measured value lower limit exceeded is triggered.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
-2000 mV