Lower hysteresis
Lower hysteresis of the dosing rate. If the dosing rate exceeds the lower
limit by this hysteresis value, then the event Dosing rate lower limit ok
is triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
0.20 mL/min
Upper limit
Upper limit of the dosing rate. If the dosing rate exceeds this limit, the
event Dosing rate upper limit exceeded is triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
166.00 mL/min
Titrino
Range
0.00 ... 150.00 mL/min
Default value
150.00 mL/min
Upper hysteresis
Upper hysteresis of the dosing rate. If the dosing rate once again falls
below the upper limit by this hysteresis value, then the event Dosing rate
upper limit ok is triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
0.20 mL/min
Action
Selection of the event to be triggered if the upper or lower limit is excee-
ded:
Selection
Cancel determination | Cancel command
(Titrando only) | Wait for [Continue] | Wait for
limit ok | none
Default value
none
Cancel determination
The running STAT command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando only)
The running STAT command will be canceled, then the next com-
mand will be carried out.