2 external measured value – Metrohm tiamo 2.3 (process analysis) User Manual
Page 945
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
929
Selection
empty | ppm | % | g/L | mg/L | mg/mL | mg/
100g | mol/L | mmol/L | mL | g | mg | µg | °C | µL
| s | mS/cm
Default value
empty
Decimal places
Number of decimal places with which the calculated measured value will
be displayed.
Range
0 ... 5
Default value
2
Comment
Comment on the calculated measured value.
Input
250 characters
Default value
'empty'
5.6.3.11.2
External measured value
Dialog window: Method
▶ DET/MET/SET/KFT/KFC/STAT/MEAS/DOS ▶ Proper-
ties...
▶ Additional measured values ▶ [New]/[Properties] ▶ External measured
value
With the option Additional external measured values, a maximum of
3 new measured values can be applied and saved from other existing
measuring commands (e.g. MEAS) which are running at the same time in
other tracks. These new measured values can also be displayed in curves
under the designation Extern 1…3 and used in formulas as variables
*.EX1…3.
Command
Selection of the command from which the additional measured value will
be taken over.
Selection
Selection from existing measuring commands
Response
Selection of the quantity from the command which will be adopted as
quantity of the additional external measured value.
Selection
Selection from existing measuring quantities
Unit
Display of the unit of the external measured value.
Comment
Comment on the external measured value.