2 kft ipol - general/hardware, General/hardware – Metrohm tiamo 2.3 (process analysis) User Manual
Page 801
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
785
Designation
Description
.MI.MEA
Minimum measured value in the unit of the measured
value
.MI.TEM
Temperature with minimum measured value in °C
.MI.TIM
Time for the minimum measured value in s
.MI.VOL
Volume at minimum measured value in mL
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in measuring point list
.SME
Start measured value (measured value after processing
the start conditions) in the unit of the measured value
.STE
Start temperature (temperature after processing of the
start conditions) in °C
.STY
Type of stop with which the command was stopped: 1
= normal; 0 = manual or after error
.SVA
Start volume absolute (volume that was added accord-
ing to the start condition "start volume") in mL
.SVT
Total start volume (volume that was added in depend-
ence on all three start conditions) in mL
.TITER
Titer value of the solution used for the command
5.6.3.5.2.2
KFT Ipol - General/Hardware
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the control device, the dosing device, the sen-
sor and the stirrer are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined