Metrohm tiamo 2.3 (process analysis) User Manual
Page 903
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
887
■
Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.
Command variables
The following command variables are generated in the method run by the
command STAT U and can be used in formulas under the designation
'Command name.Variable designation':
Designation
Description
.BLV
Blank value of the sensor used for the command (only
for ISE sensors)
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used
for the command
.DBL
Total duration for the processing of the command in s
.DSC
Time for processing all start conditions s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode slope of the sensor used for the command (in
mV for ISE sensors)
.ETE
End temperature (temperature following processing of
the command) in °C
.EVT
End volume (total dosed volume at the end of the
command) in mL
.FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) = Com-
mand has never ended
.FP{x}.MEA
Measured value for the fixed endpoint × (1…9) in mV
.FP{x}.TEM
Temperature for the fixed endpoint × (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1…9) is reached
.FP{x}.VOL
Volume for the fixed endpoint × (1…9) in mL