5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
570
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Designation
Description
.STY
Type of stop with which the command was stopped: 1
= normal; 0 = manual or after error
.SVA
Start volume absolute (volume that was added accord-
ing to the start condition "start volume") in mL
.SVM
Start volume measured value (volume which was
added according to the start measured value given in
the start conditions) in mL
.SVS
Start volume slope (volume that was added according
to the start condition "start slope") in mL
.SVT
Total start volume (volume that was added in depend-
ence on all three start conditions) in mL
.TITER
Titer value of the solution used for the command
5.6.3.3.2.2
MET pH - General/Hardware
Tab: Method
▶ MET pH ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the control device, the dosing device, the sen-
sor and the stirrer are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead