10 meas cond, 1 meas cond - overview, Meas cond – Metrohm tiamo 2.3 (process analysis) User Manual
Page 1027
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1011
Additional external measured values
Additional external measured values
on | off (Default value: off)
If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-
[New]
Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 929).
[Properties]
Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 929).
[Delete]
Delete the external measured value selected in the table.
5.6.4.10
MEAS Cond
5.6.4.10.1
MEAS Cond - Overview
Dialog window: Method
▶ MEAS Cond ▶ Properties... ▶ MEAS Cond - 'Command
name'
Command for Conductivity measurements.
Instruments
This command can be executed with the following devices:
Conductometer: 712, 856
Appearance
The command has the following appearance:
Parameters
The parameters for the command MEAS Cond are configured on the fol-
lowing 4 tabs:
■
General/Hardware
Parameters for devices, sensors and stirrers.