8 met cond - additional measured values, Additional measured values – Metrohm tiamo 2.3 (process analysis) User Manual
Page 704
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
688
■■■■■■■■
tiamo 2.3 (for Process Analysis)
Window
A range (window) can be defined on the measured value axis, on the vol-
ume axis or on the time axis. The break point evaluation will only be car-
ried out in the defined window. Only the first break point in the defined
window will be recognized.
Selection
Measured value | Volume | Time | off
Default value
off
Lower limit
Value for the lower limit of the window.
Window = Measured value
Range
-2000.0 ... 2000.0 mS/cm
Default value
-2000.0 mS/cm
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.
Window = Measured value
Range
-2000.0 ... 2000.0 mS/cm
Default value
2000 mS/cm
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
Window = Time
Range
0 ... 999999 s
Default value
999999 s
5.6.3.3.6.8
MET Cond - Additional measured values
Tab: Method
▶ MET Cond ▶ Properties… ▶ Additional measured values
Command name
Name of the command.
Input
25 characters