5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
716
■■■■■■■■
tiamo 2.3 (for Process Analysis)
■
Titration parameters
Parameters for the run of the titration.
■
Stop conditions
Definitions of conditions which cause the stop of the titration.
■
Conditioning
Enable/Disable the conditioning. Definition of the start drift, drift cor-
rection and stop conditions.
■
Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.
■
Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.
Command variables
The following command variables are generated in the method run by the
command SET U and can be used in formulas under the designation
'Command name.Variable designation':
Designation
Description
.BLV
Blank value of the sensor used for the command (only
for ISE sensors)
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.CBY
Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated
.COK
Command status; 1 = Condition requirement fulfil-
led, 0 = Condition requirement not fulfilled
.CONC
Concentration of the solution used for the command
.CYL
Cylinder volume of the exchange or dosing unit used
for the command
.DBL
Total duration for the processing of the command in s
.DRI
Current and/or last drift for drift correction in µL/min
.DSC
Time for processing all start conditions s
.DTI
Time for drift correction (time from the start of the
titration to the end of the command) in s
.EME
End measured value (measured value after processing
of the command) in the unit of the measured value