beautypg.com

Delta DVP-ES2 User Manual

Page 357

background image

3 . I n s t r u c t i o n S e t

3 - 2 4 1

Points to note:

1.

There is no limitation on the times of using this instruction. However, the register No.

designated in S

3

~ S

3

+19

cannot be repeated.

2.

For 16-bit instruction, S

3

occupies 20 registers. In the program example above, the area

designated in S

3

is D100 ~ D119.

3.

Before the execution of PID instruction, users have to transmit the parameters to the

designated register area by MOV instruction. If the designated registers are latched, use

MOVP instruction to transmit all parameters only once

4. Settings

of

S

3

in the 16-bit instruction:

Device

No.

Function

Setup Range

Explanation

S

3

:

Sampling time (T

S

)

1~2,000

(unit: 10ms)

Time interval between PID

calculations and updates of MV. If T

S

= 0, PID instruction will not be

enabled. If T

S

is less than 1 program

scan time, PID instruction sets S

3

as 1

program scan time, i.e. the minimum

T

S

has to be longer than the program

scan time.

S

3

+1:

Propotional gain

(K

P

)

0~30,000(%)

The proportion for

magnifying/minifying the error

between SV and PV.

Integral gain (K

I

) 0~30,000(%)

The proportion for

magnifying/minifying the integral

value (The accumulated error). For

control mode K0~K5.

S

3

+2:

Integral time

constant (T

I

)

0~30,000 (ms)

For control mode K10

Derivative gain (K

D

)

-30,000~30,000

(%)

The proportion for

magnifying/minifying the derivative

value (The rate of change of the

process error). For control mode

K0~K5

S

3

+3

:

Derivative time

constant (T

D

)

-30,000~30,000

(ms)

For control mode K10

S

3

+4:

Control mode

0: Automatic control

1: Forward control (E = SV - PV).

This manual is related to the following products: