Appendix K. VersaDoc

K-9

The following table provides recommended exposure times for various applications

Table 1: Recommended Exposure Times and Lenses

Sample

Recommended

Exposure

Lens & Filter

Accessories Used

Fluorescent Stain Gel

3–30 sec.

Zoom/IR

None

Fluorescence End-
Label Gel

30 sec.–5 min.

Zoom/IR

None

Fluorescent Blot

0.5–5 sec.

Zoom/IR

Sample/Chemi
Tray (if sample is
small)

Chemifluorescent
Blot

0.5–5 sec.

Zoom/IR

None

Colorimetric Gel

0.1–1 sec.

Zoom/IR

White light con-
version screen

Colorimetric Blot

0.1–1 sec.

Zoom/IR

Sample/Chemi
Tray (if sample is
small)

X-ray film

0.1–1 sec.

Zoom/IR

White light con-
version screen

Weak

Chemiluminescence

1

5–10 min.

50 mm

Sample/Chemi
Tray

(if sample is
small)

Strong

Chemiluminescence

1

10 sec.–2 min.

50 mm

Sample/Chemi
Tray

(if sample is
small)

1

For chemi applications, the 50mm lens is recommended. Always remove the 660 fil-

ter.