B differential display searches – Bio-Rad Quantity One 1-D Analysis Software User Manual

Page 205

background image

Quantity One User Guide

9-2

See section 6.2.d, Normalizing for Quantity, for more information on normalization.

9.1.b Differential Display Searches

After you have normalized the gel, select Differential Display from the Analysis
menu. A dialog will open in which you can specify certain parameters.

Fig. 9-1. The Differential Display dialog box.

You can perform two types of searches on bands across a differential display gel: an
Outliers search or a Trends search. Both are determined by the Percentage and
Quantity values that you specify in the Differs By field.

Outliers

Select the Outliers button to highlight bands with a normalized quantity that differs
from the mean normalized quantity for that band type by the specified percentage
and/or the specified quantity. An outlier is defined as a band with normalized quantity
(n

q

) that satisfies one of the following criteria:

1.

n

q

> ratio x mean and

n

q

> mean + quantity

2.

n

q

< (1/ratio) x mean and

n

q

< mean – difference

Where: ratio = 1.0 + (percentage/100)