Index
Error Messages and Their Corrective
Actions
Software Configurator
Warning
Code
Warning message
What to do
W3331
The µR1000 and µR1800 cannot be distinguished
automatically. Please select the correct recorder model.
Select µR1800 or µR1000 manually.
W3332
The recorder model can’t be recognized. It will be
regarded as not connected.
Check the model of the connected recorder.
W3333
The VR100 and VR200 cannot be distinguished
automatically. Please select the correct recorder model.
Select VR100 or VR200 manually.
Message
Code
Message
What to do
M3361
Model determination was successful.
— — — —
Hardware Configurator
Code
Error message
What to do
E3401
Failed to send command. Settings don’t fit currently
connected model or computations may be in progress.
Check the model of the connected recorder. If
computation processing is in progress, stop the
computation then send command.
E3402
Failed to receive command. Settings don’t fit currently
connected model.
Check the model of the connected recorder.
E3403
File is invalid. Hardware Configuration settings will
return to default.
Check if you have specified a hardware
configuration setting file.
E3404
Cannot send command while IC memory is in use.
Stop the IC memory operation, then send command.
E3405
Data memory is full. Please delete all data and send
again.
Clear the data memory, then send data.
E3408
Invalid user name or password.
Use the Software Configurator to check the user
name and password.
E3409
Data cannot be send when logged in as User. Login as
administrator and send again.
Use the Software Configurator to set the login name
of administrator.
E3410
Some setting could not be sent.
Check the model and options of the connected
recorder.
E3411
Cannot send command while computations or memory
sampling are in progress.
Stop the recorder’s memory sampling or computation
processing, then send command.
E3412
This file does not fit settings made with DAQLOGGER
Software Configurator. Cannot read this file.
Abort loading. Check the contents of the file.
E3413
Communication not possible. Temperature units don't
match.
Check the temperature unit.
E3414
Connected to different model.
Check the connected device.
E3415
Please change connected hardware to SET mode.
Check the connected device.
E3416
Please enter an equation.
Set an expression.
E3417
Unknown symbol is entered. Please change equation.
Check the setup parameters.
E3418
Invalid channel number is entered. Please change
equation.
Check the setup parameters.
E3419
Incorrect use of parenthesis. Please change equation.
Check the setup parameters.
E3420
Grammatical error. Please change equation.
Check the setup parameters.
E3421
These data were generated by another model.
Check the setup parameters.
E3422
Temperature units in file must match software.
Check the temperature unit.
E3423
Failed to load data. Some data initialized.
Check the contents of the file.
E3424
Invalid channels are included in the copy source.
Check the copy source and copy destination
settings.
E3425
Character string is incorrect.
Check the setup parameters.
E3426
Invalid channel number.
Check the channel number.
E3427
For input range, please keep CH numbers Left < Right.
Check the setup parameters.
E3428
Channel number is already being used!
Check the setup parameters.
E3429
Incorrect file name!
Check the file name and extension.
E3430
Now Controlling. Can’t store settings.
Send the settings after halting control operations on
the CX.
9.1 Error Messages and Their Corrective Actions