Index, Display – Yokogawa DAQWORX User Manual
Page 237

7-61
IM WX101-01E
1
2
3
4
5
6
7
8
9
Index
Setting the Recorder
Display
Trend interval [/div]
Select the time corresponding to 1 division of the time axis on the trend display from
below: You cannot specify a trend interval that is faster than the scan interval. See the
table under “Save Interval” below.
15s*, 30s, 1min, 2min, 5min, 10min, 15min, 20min, 30min, 1h, 2h, 4h, and 10h
* Can be set on the DX1002, DX1002N, DX1004, DX1004N, DX2004 and DX2008.
Save Interval (when recording display data)
Select the size of a record data file. The recorded data is divided by the file size
specified here. The available settings vary depending on the Trend interval setting.
Trend interval
Selectable range of
auto save interval
Trend interval
Selectable range of
auto save interval
Trend interval
Selectable range of
auto save interval
*1 Selectable on the DX1002, DX1002N, DX1004, DX1004N, DX2004, and DX2008 (release
number 3 or later).
*2 Selectable in fast sampling mode on the DX1006, DX1006N, DX1012, DX1012N, DX2010,
DX2020, DX2030, DX2040, and DX2048 (release number 3 or later).
5 s*
1
10 s*
1
15 s*
2
30 s
1 min
10 min to
10 min to
10 min to
10 min to
10 min to
12 hours
1 day
3 days
7 days
14 days
2 min
5 min
10 min
15 min
20 min
10 min to
10 min to
10 min to
10 min to
1 hour to
14 days
31 days
31 days
31 days
31 days
30 min
1 h
2 h
4 h
10 h
1 hour to
1 hour to
2 hours to
4 hours to 8 hours to
31 days
31 days
31 days
31 days
31 days
Circular Time Per revolution [/rev]
Select the time of revolution from [20min]* to [4week].
* For release number 2 or earlier, this can only be specified on the DX2004 and DX2008.