Setting the Recorder
• WR_TRIG : Saves measured/computed data on the built-in RAM disk.
• RD_TRIG : Reads measured/computed data from the built-in RAM disk.
• LD_TRG1-3 : Reads setup data from the built-in RAM disk.
DC100
• DATA_WR : Writes the data for one scan (one data/channel) retained at the
generation of an event. To execute this, it is necessary to set LOGIC to the writing
period.
• TRIG : Generates a trigger to start writing. To execute this, it is necessary to set
TRIG to WRITE TRIG.
• START : Opens a file and starts the writing operation. It is the same function as the
START key on the operation panel.
• PAUSE : In the case of edge action, writing is suspended at the generation of
an event. In the case of level action, by setting the alarm, remote, relay or internal
switch to the event, writing is suspended at the detection of an event and resumes at
the resetting of the event (see the following figure). Setting the timer, match time or
manual function key to the event repeats suspension and resumption of data writing at
the generation of each event.
• CP_FDD : All files in the built-in RAM disk are copied on a floppy disk. Whether to
cancel files after copying or not depends on the setting in the MEM SET of the SET
UP menu. Copying after a conversion to ASCII format cannot be done.
• CP_SCS0 to CP_SCS7 :
All files in the built-in RAM disk are copied on a MO, ZIP or PD. Whether to cancel
files after copying or not depends on the setting in the MEM SET of the SET UP menu.
Copying after a conversion to ASCII format cannot be done.
Displayed only on models with the C/5 option. SCS0 to SCS7 represent the SCSI ID
numbers. Check the ID numbers of SCSI devices that are connected, because ID
numbers are displayed even if there are no devices connected to them.
FLAG (only for Math option models; DC style no. 7+, DA100/DR style no 8 or higher)
• F01 to 16 : Set the flag number (F01 to F16) to 1. F01 to F16 are normally 0. Flags
can be placed in computing equations to hold the computed result using a certain
event as a trigger, or reset to 0.
FLOPY
• LD_TRG1-3 : Reads setup data from the floppy disk.
SCSI0 to SCSI7 (for DC100 with /C5 option only)
• LD_TRG1 to 3 : Read setup data from the SCSI device (MO, ZIP or PD).
SCSI0 to SCSI7 represent the SCSI ID numbers. Check the ID numbers of SCSI
devices that are connected, because ID numbers are displayed even if there are no
devices connected to them.
REPORT (available if the instrument is equipped with the REPORT function.)
• START (Edge action) : Starts making up a report.
• STOP (Edge action) : Stops making up a report.
(No Start/Stop for Level action)
Basic Setup for hourly, daily, or monthly report function must be ON.
• REPORT (Level action) :
Starts/stops making up a report.
Note
If, in the SETUP mode, the report function is set to OFF for all data items, you cannot select the
action item REPORT.