beautypg.com

Rapidlab 800 maintenance schedule month/year – Bayer HealthCare Rapidlab 800 User Manual

Page 662

background image

Rev. E

Monthly Maintenance

1

2

3

4

5

6

7

8

9

10

11

12

Exchange the cleaning solutions

Replace 7.3 buffer installed for

30 days

Replace Cal G/L reagent installed for

30 days

Inspect the capillary seal

Bimonthly Maintenance

1

2

3

4

5

6

Replace 6.8 buffer and Wash/Zero

reagents installed for 60 days

Replace sample tubing

Quarterly Maintenance

1

2

3

4

Clean the sample chamber

Clean the hemolyzer

RAPIDLAB 800 MAINTENANCE SCHEDULE Month/Year ________________