D76 coćox electronics error, Problem, Solutions – Bayer HealthCare Rapidlab 800 User Manual
Page 332

4Ć48
800 Series Operator's Manual
D76 COĆox Electronics Error:
Problem
The system detects an error in the electronics system.
Qualifiers:
1 2
3
4
5
6
7
8
9
10 11 12
Solutions
If the qualifier is . . .
Then . . .
1, 2, 3, 5, 8, 11
perform the CO-ox Lamp On/Off test, and then perform a
one-point calibration.
Shut the system down and restart:
1.
From the Menu screen, select
7 System Utilities
and
press
Enter
.
2.
Select
3 Shutdown
and press
Enter
.
3.
Press
Yes
.
4.
Wait at least 1 minute and then disconnect the power cord
from the power source.
5.
Wait at least 10 seconds and then connect the power cord
to the power source.
4, 7, 9, 12
shut the system down and restart.
6
check the temperature control as described in
Temperature/pAtm Test, page 4-69.
Shut the system down and restart.
10
perform the Pump Functions Test as described in Pump
Functions Test, page 4-58.
If the D code reappears, contact your Service Representative.