Recalling calibration data – Bayer HealthCare Rapidlab 800 User Manual
Page 159

2Ć61
Operating the System
Recalling Calibration Data
You can print calibration data and send the data to an LIS or a data management
system. You cannot edit calibration data.
1.
Access the Recall Calibration Data screen from the Menu screen:
a. Select
4 Data Recall
and press
Enter
.
b. Select
4 Calibration Data
and press
Enter
.
2.
Select a calibration type and press
Enter
.
3.
Type the search criteria and press
Enter
.
NOTE:ĄĂ
The Calibration Date fields contain the current day’s date. If you do
not complete any other fields, the system recalls all one-point calibrations for
this day.
If you want to search . . .
Then . . .
for a selected date
type the date in both Calibration Date fields.
from the earliest date to a
specific date
leave the Calibration Date From field blank and type
the specific date in the Calibration Date To field.
from a specific date to the
present date
type the start date in the Calibration Date From field
and leave the Calibration Date To field blank.
by shift or any other
specific time period
type the start time in the Calibration Time From field
and type the end time in the Calibration Time To
field.
4.
Press
Done
.
If . . .
Then . . .
more than one calibration
is found
the Calibration Data Search Log screen appears as
shown in Figure 2-34. Continue with step 5.
one calibration is found
the Calibration Data Search Result screen appears, as
shown in Figure 2-35. Continue with step 6.
4
Menu Code
4