Bayer HealthCare Rapidlab 800 User Manual
Page 251

3Ć79
Maintaining the System
Figure 3Ć53ā.ăFilling the Electrode Compartment
Electrode
Compartment
OĆring
Sample Path
4M KCl Fill Solution
OĆring
Reference Sensor Cassette
(KCI Reservoir)
CAUTION:
Do not touch the internal electrode wire. The wire is fragile and is
easily damaged.
8.
Use the hex tool to remove the new internal electrode from its container.
9.
Insert the internal electrode into the compartment and screw it into place with
the hex tool, ensuring that you do not cross-thread the electrode.
10. Fill the KCl reservoir in the sensor:
a. Remove the reservoir cap with the hex tool and set it aside as shown in
Figure 3-54.
Figure 3Ć54.āăRemoving the Reservoir Cap with the Hex Tool
Hex Tool
Reservoir Cap
OĆring
OĆring
Sample Path
KCI Reservoir
Vent Hole
b. Partially fill the KCl reservoir by gently squeezing the KCl fill solution
container as shown in Figure 3-55.
!