Performing the Automatic Clean Sequence

Use this procedure to manually initiate the cleaning sequence. The system
automatically initiates the cleaning sequence every 24 hours at 02:00.

1.

Access Auto Clean from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

6 Auto Clean

and press

Enter

.

NOTE:ĄĂ

To interrupt the Auto Clean sequence at any time, press

Cancel

.

The Auto Clean sequence takes 10 minutes. A message box appears prompting
you to perform a two-point calibration.

2.

Press

Yes

to perform a two-point calibration.

Performing a Wash

1.

Access wash from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

8 Wash

and press

Enter

.

2.

Press

Home

to return to the Ready screen.

Priming the System

1.

Access the Prime screen from the Menu screen:

a. Select

2 Maintenance

and press

Enter

.

b. Select

3 Prime

and press

Enter

.

2.

Select the appropriate reagent to prime and press

Enter

.

3.

Press

Done

.

4.

Press

Home

or

Menu

when the Prime menu appears.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

5.

Take the appropriate action.

If . . .

Then . . .

you want to perform a
two-point calibration

press

Yes

.

you do not want to initiate a
two-point calibration

press

No

.

press

Home

to return to the Ready screen.