Performing the automatic clean sequence, Performing a wash, Priming the system – Bayer HealthCare Rapidlab 800 User Manual
Page 232

3Ć62
Rapidlab 800 Operator's Manual
Performing the Automatic Clean Sequence
Use this procedure to manually initiate the cleaning sequence. The system
automatically initiates the cleaning sequence every 24 hours at 02:00.
1.
Access Auto Clean from the Menu screen:
a. Select
2 Maintenance
and press
Enter
.
b. Select
6 Auto Clean
and press
Enter
.
NOTE:ĄĂ
To interrupt the Auto Clean sequence at any time, press
Cancel
.
The Auto Clean sequence takes 10 minutes. A message box appears prompting
you to perform a two-point calibration.
2.
Press
Yes
to perform a two-point calibration.
Performing a Wash
1.
Access wash from the Menu screen:
a. Select
2 Maintenance
and press
Enter
.
b. Select
8 Wash
and press
Enter
.
2.
Press
Home
to return to the Ready screen.
Priming the System
1.
Access the Prime screen from the Menu screen:
a. Select
2 Maintenance
and press
Enter
.
b. Select
3 Prime
and press
Enter
.
2.
Select the appropriate reagent to prime and press
Enter
.
3.
Press
Done
.
4.
Press
Home
or
Menu
when the Prime menu appears.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
5.
Take the appropriate action.
If . . .
Then . . .
you want to perform a
two-point calibration
press
Yes
.
you do not want to initiate a
two-point calibration
press
No
.
press
Home
to return to the Ready screen.
2
Menu Code
6
2
Menu Code
8
2
Menu Code
3