Analyzing syringe samples – Bayer HealthCare Rapidlab 800 User Manual
Page 104

2Ć6
Rapidlab 800 Operator's Manual
Analyzing Syringe Samples
The following tables list the minimum volumes required by each system to analyze
syringe samples:
Table 2Ć1.āăMinimum Sample Volumes
System
Minimum Sample Volume
840
844, 845
0
90
µ
L
140
µ
L
850
854, 855
110
µ
L
160
µ
L
860
864, 865
125
µ
L
175
µ
L
The system automatically aspirates the sample. If you want the system to analyze a
different set of parameters for the sample, press Change Parameters to select
another panel. Refer to Selecting Parameter Panels, page 2-29, for more
information.
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1.
Prepare the sample and insert the syringe into the sample port, as shown in
Figure 2-2.
Figure 2Ć2.ā Inserting a Syringe Sample
Sample Port
Syringe Sample
CAUTION:ĄĂ
To ensure the accuracy of the CO-ox measurement, close the
CO-ox cover before pressing Analyze.
!