D19 fluid detector error – Bayer HealthCare Rapidlab 800 User Manual
Page 314
4Ć30
800 Series Operator's Manual
D19 Fluid Detector Error:
Problem
Fluid detector millivolt (mV) reading is beyond predefined limits.
Qualifiers:
FD1
FD1A
FD2
Solutions
Check the sample path in the measurement module for obstructions or leaks.
Ensure that O-rings are in place on the sensors and that the sensors are aligned
correctly as described in Removing and Checking the Sensors on page 4-65.
If you see obstructions, remove them as described in Removing Obstructions
from the Measurement Module, page 4-82.
Perform the Fluid Detector Test as described in Fluid Detector Test, page 4-66.
Deproteinize the sample path as described in Deproteinizing the Sample Path in
Section 3, and repeat the Fluid Detector Test.
If the D code reappears, contact your Service Representative.
D19