Performing a barometer calibration – Bayer HealthCare Rapidlab 800 User Manual
Page 157

2Ć59
Operating the System
Performing a Barometer Calibration
Use this procedure to calibrate the internal atmospheric pressure sensor to a
barometer in your laboratory.
1.
Access the Barometer Calibration screen from the Menu screen:
a. Select
1 Calibration
and press
Enter
.
b. Select
8 Barometer
and press
Enter
.
2.
Type the correct atmospheric pressure and press
Enter
.
The acceptable range for atmospheric pressure is 400 – 825 mm Hg
(53.0 – 110.0 k Pa).
3.
Press
Done
to save the new atmospheric pressure and return to the Ready
screen.
1
Menu Code
8